However, compared to wear particles from PEEK and UHMWPE, PEEK-OPTIMA particulates exhibited less cytotoxicity and demonstrated a size-dependent cytokine release with an increasing trend toward smaller particles (10–2 μm). 198 A study on PEEK-on-ceramic cervical artificial discs (subjected to 10 million cycles using a spine simulator) revealed that PEEK endplates were the primary source of wear ranging from 0.9 ± 0.2 to 2.8 ± 0.6 mg/million cycles for low and high number of cycles, respectively.
← all excerpts
Spinal implant wear particles: Generation, characterization, biological impacts, and future considerations.