Estimation of Photon Specific Absorbed Fractions in Digimouse Voxel Phantom using Monte Carlo Simulation Code FLUKA.

J Biomed Phys Eng · 2016 · PMC5219571 · PMID 28144589

1
hedged sentence
closest p
boldest claim

The sentences

a decreasing trendno p-value reported
It can be concluded that, bladder and testis also have the smallest SAF values because they are far from brain. Primary Photon Energy As shown in Figure 2 (SAF values for self-irradiation in liver, stomach, heart, kidney and pancreas) and Figure 3 (SAF values for cross-irradiation in stomach, pancreas, spleen, heart and kidney with liver as source organ), photon SAF values for self-irradiation and cross-irradiation show a decreasing trend with increasing energy in the photon energy range of 15 keV to 100 keV and 500 keV to 4000 keV.