Quantification of Wear and Deformation in Different Configurations of Polyethylene Acetabular Cups Using Micro X-ray Computed Tomography.

Materials (Basel) · 2017 · PMC5503355 · PMID 28772616

1
hedged sentence
closest p
boldest claim

The sentences

an increasing trendno p-value reported
The XLPE_VE acetabular cups, under HDA waveforms, showed an increasing trend for mass loss: more than 3 times with respect to the other two configurations ( Table 2 ). μ-CT investigations showed that different wear and deformations can be achieved when different materials and/or different types of wear tests are considered, proving that a mere weight loss assessment is not sufficient for a complete understanding of components’ behavior.