Performance of Wearable Pulse Oximetry During Controlled Hypoxia Induction: Instrument Validation Study.

JMIR Form Res · 2026 · PMC13026435 · PMID 41894520

2
hedged sentences
closest p
boldest claim

The sentences

a decreasing trendno p-value reported
A rms across 5% SaO 2 bins demonstrated a decreasing trend ( Figure 4 ) for both devices with increasing SaO 2 .

also in 47,180 other papers

showed a trendno p-value reported
There are three main findings of our study: (1) both the Masimo MightySat Rx and the Apple Watch Series 7 often overestimate true blood oxygen saturation values, with 44% and 56% of their measurements in this study, respectively, found to be higher than the ground truth SaO 2 measures; (2) both devices have lower accuracy during hypoxemia than at normal blood oxygen saturation; and (3) both devices showed a trend toward lower accuracy when the ODR is larger than 2% per minute, but the difference is not statistically significant.