Impact of real-time continuous glucose monitoring on glucose variability in critically ill patients

Crit Care · 2010 · PMC2934457

1
hedged sentence
0.0870
closest p · 1.7× alpha
0.0870
boldest claim

The sentences

showed a trendP = 0.087so close (0.05 < p ≤ 0.1)
Although the mean sensor glucose during the first 24 hours in the ICU showed a trend towards lower values in the intervention group (real-time CGM vs control: Glu MEAN1 : 104.40 ± 2.00 vs 109.23 ± 1.96 mg/dl; P = 0.087), variability of sensor glucose during the first 24 hours did not differ significantly (real-time CGM vs control: Glu SD1 : 20.85 vs 18.46 mg/dl; P = 0.201).