On the use of a convolution-superposition algorithm for plan checking in lung stereotactic body radiation therapy.

J Appl Clin Med Phys · 2016 · PMC5874108 · PMID 27685114

1
hedged sentence
closest p
boldest claim

The sentences

clearly significantno p-value reported
Using the recommended ICRU 83 dose reporting parameters, there were clearly significant differences ( > 5 % ) between the M3D and TPS (XiO and Eclipse AAA) calculations that would be difficult to resolve by direct measurement for the reasons previously discussed.