Factors Affecting Repeatability of Foveal Avascular Zone Measurement Using Optical Coherence Tomography Angiography in Pathologic Eyes.

Clin Ophthalmol · 2020 · PMC7147611 · PMID 32308362

2
hedged sentences
closest p
boldest claim

The sentences

Among eyes with macular edema, the ICC for the right and left eyes were 0.775 and 0.950, respectively [ Table 3 ], demonstrating a trend toward a decrease in repeatability of FAZ measurements using machines from different vendors in patients with macular edema in the right eye which did not reach statistical significance based on 95% confidence intervals.

also in 111,027 other papers

an overall trendno p-value reported
Finally, there was an overall trend for the right eye to have lower ICC for inter-instrument repeatability of FAZ measurement when compared to the left eye.