The ML8 required numerically higher pressures (0.71 ± 0.15 N); however, this did not reach statistical significance ( p = .08). 4 DISCUSSION In this bench study of the 3.0‐mm Magmaris BRS compared with previously tested scaffolds and stents, the main findings were as follows: MB postdilatation with balloon diameters less than 4.4 mm did not cause fracture.
← all excerpts
Mechanical properties of the drug-eluting bioresorbable magnesium scaffold compared with polymeric scaffolds and a permanent metallic drug-eluting stent.