Accuracy and Reliability of Whole Blood Bilirubin Measurements Using a Roche Blood Gas Analyzer for Neonatal Hyperbilirubinemia Screening and Risk Stratification.

Front Pediatr · 2022 · PMC9289128 · PMID 35859945

1
hedged sentence
closest p
boldest claim

The sentences

showed a trendno p-value reported
Consistent with the results of the Passing–Bablok regression analysis, the Bland–Altman plot ( Figure 2 ) showed a trend of overestimation with Roche TBiL at low bilirubin levels, with eight paired data points (8/309) falling below the lower 95% LoA.